# This is calibration data for the aerosol Raman spectrometer
# Used in the Doughty and Hill JQSRT papers (2019)
# The calibration is for 8/25/2016
# Calibration values are a combination of the two full
# calibrations measured before and after this day.
# full calibrations done using the lines from a neon lamp.
# The determined laser wavelengths measured using
# the 520.7 cm-1 peak on a silicon slab are an average
# of eight silicon measurements made during the full
# calibrations and during field calibrations with only
# the silicon slab used.
# Uncertainty notes: Between 400 and 1800 cm-1, its +0.56/-0.86,
# above 2700, its +3.29/-3.58. These are the worst case errors. real error is likely to be less. Errors change with
# wavelength, with the lowest magnitudes between 500-1500 cm-1
# but some cross zero at two places in this range.
# Please see doughty and hill (2019) in the supplement for a
# comparison of different calibrations
#
Spectrometer Coefficients (ax^3 + bx^2 + cx + d):-1.2483276e-09,-6.87657289e-06,0.171035657,641.717874
Y intercept:642.986773071