MET/CAL Procedure
=============================================================================
INSTRUMENT: Main
DATE: 2024-03-05 10:20:07
AUTHOR: Dan Erlandsen
REVISION:
ADJUSTMENT THRESHOLD: 70%
NUMBER OF TESTS: 1
NUMBER OF LINES: 104
=============================================================================
STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON
1.001 DISP Calibration procedure for Druck DPI 610
# 1.002 CALL milliAmps
# 1.003 CALL Voltage
# 1.005 CALL Vacuum
1.002 DISP Druck DPI 610 V in til fluke 8558A
1.003 MATH M[3] = 30
1.004 HEAD {V DC Measure}
1.005 HEAD -2{[M3] V Range}
1.006 MATH M[1]= 0.5
1.007 CALL Voltage
1.008 MATH M[1]= 1
1.009 CALL Voltage
1.010 MATH M[1]= 2
1.011 CALL Voltage
1.012 MATH M[1]= 5
1.013 CALL Voltage
1.014 MATH M[1]= 10
1.015 CALL Voltage
1.016 MATH M[1]= 20
1.017 CALL Voltage
1.018 MATH M[1]= 30
1.019 CALL Voltage
1.020 DISP Koble DPI 610 til 5550A
1.021 DISP Sett DPI 610 til mA measure (I)
1.022 HEAD {mA DC Measure}
1.023 HEAD -2{55 mA Range}
1.024 MATH M[1]= 1
1.025 CALL MilliAmps
1.026 MATH M[1]= 4
1.027 CALL MilliAmps
1.028 MATH M[1]= 12
1.029 CALL MilliAmps
1.030 MATH M[1]= 20
1.031 CALL MilliAmps
1.032 MATH M[1]= 30
1.033 CALL MilliAmps
1.034 MATH M[1]= 55
1.035 CALL MilliAmps
#Header and range
1.036 DISP Test procedure for pressure calibration
1.037 MATH M[3] = 7000
1.038 HEAD {Pressure measurement}
1.039 HEAD -2{[M3] mBar range}
1.040 MATH M[1] = 500
1.041 CALL Pressure
1.042 MATH M[1] = 1000
1.043 CALL Pressure
1.044 MATH M[1] = 3000
1.045 CALL Pressure
1.046 MATH M[1] = 5000
1.047 CALL Pressure
1.048 MATH M[1] = 7000
1.049 CALL Pressure
1.050 DISP Test procedure for pressure calibration
1.051 MATH M[3] = 7000
1.052 HEAD {Pressure measurement}
1.053 HEAD -2{[M3] mBar range}
1.054 MATH M[1] = 0
1.055 CALL Pressure
1.056 MATH M[1] = -300
1.057 CALL Pressure
1.058 MATH M[1] = -600
1.059 CALL Pressure
1.060 MATH M[1] = -800
1.061 CALL Pressure