MET/CAL Procedure
=============================================================================
INSTRUMENT: Pressure
DATE: 2024-08-29 09:00:15
AUTHOR: Dan Erlandsen
REVISION:
ADJUSTMENT THRESHOLD: 70%
NUMBER OF TESTS: 2
NUMBER OF LINES: 58
CONFIGURATION: Fluke 8558A
=============================================================================
STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON
1.001 ASK- R N M
1.002 8558A +Current = 24mA; Setup
#Tell the fluke its setpoint, then set it to control mode. 10 is its current GPIB address
1.003 IEEE [@10]
1.004 IF (M[1] == 0)
#IF GAUGE and set point is 0, vent to atm. Else, if mode is ABS and setpoint is 0, set mode to control
1.005 IF (M[21] == 1)
1.006 IEEE OUTP:MODE vent
1.007 ELSE
1.008 IEEE SOUR:PRES [M1]
1.009 IEEE OUTP:MODE control
1.010 ENDIF
1.011 ELSE
1.012 IEEE SOUR:PRES [M1]
1.013 IEEE OUTP:MODE control
1.014 ENDIF
#wait for pressure to stabilize
1.015 WAIT -t [M9] Waiting for stabilization
# 1.011 DISP Wait until fluke 6270A displays the READY flag, then press "advance" or enter
#Ask Fluke for its current pressure reading, then store it in MEM1 register
1.016 IEEE MEAS?[I]
## 1.007 RSLT =[MEM] mBar
1.017 MATH M[11] = MEM
#Calulate corresponding mA value
1.018 MATH MEM1 = ((MEM - M[7]) / (M[8] - M[7]) * (M[6] - M[5]) + M[5])
#Read current mA value
1.019 8558A +Current = [MEM1]mA;AverageCount = 4; Read
1.020 VSET TDESC = [M11] [S14]
1.021 MEMCX 20 mA TOL
#stop regulating pressure, since test is finished
2.001 IF (M[1] == 0)
2.002 IEEE OUTP:MODE vent
2.003 ELSE
2.004 IEEE OUTP:MODE measure
2.005 ENDIF
2.006 END