You signed in with another tab or window. Reload to refresh your session.You signed out in another tab or window. Reload to refresh your session.You switched accounts on another tab or window. Reload to refresh your session.Dismiss alert
I am using MAX30100_PulseOximeter library and is working fine. To save to chip longer life I need to stop MAX30100 and turn the LEDs off. Please advice.
The text was updated successfully, but these errors were encountered:
I believe that you can not shutdown max30100. This feature is present in max30102.
About shutdown leds... u can set current to 0.0 mA (change register with address equal 0x09), so they will be in off state.
On 6 Feb 2020, at 14:40, williamesp2015 ***@***.***> wrote:
I am using MAX30100_PulseOximeter library and is working fine. To save to chip longer life I need to stop MAX30100 and turn the LEDs off. Please advice.
—
You are receiving this because you are subscribed to this thread.
Reply to this email directly, view it on GitHub, or unsubscribe.
I am using MAX30100_PulseOximeter library and is working fine. To save to chip longer life I need to stop MAX30100 and turn the LEDs off. Please advice.
The text was updated successfully, but these errors were encountered: