Scalable Modular Ozone Generation based Sterilization Chamber ( SMOG-SC )

This project aims to lay down general design instructions for an ozone generation based sterilization chamber that could be deployed fast to medical environments. The design is layered and modular to take component availability into account.

About ozone

Ozone (O3) is an unstable gas comprising three atoms of oxygen. It is unstable because the gas will readily degrade back to its stable state, diatomic oxygen (O2) with the formation of free oxygen atoms or free radicals. The free oxygen atoms or radicals are highly reactive and they will oxidize almost anything including viruses, bacteria, organic and inorganic compounds) in contacts, making ozone an enormously powerful disinfectant and oxidizer.

⚠️ Ozone with concentration higher than 1 ppm has adverse effects on human health and the use of ozone for air disinfection is generally not recommended if people are around. ⚠️

The MAC-value is the Maximum Acceptable Concentration a human is allowed to be exposed to for a certain time and certain agent. For ozone the MAC-values is 0.06 PPM for 8 hours a day, 5 days a week (PPM = Parts Per Million). For a maximum of 15 minutes a MAC-value of 0.3 PPM is applied.

Read more: https://www.lenntech.com/library/ozone/decomposition/ozone-decomposition.htm

Temp (°C) Half live
-50 3 months
-35 18 days
-25 8 days
20 3 days
120 1,5 hours
250 1,5 seconds

Dissolved in water (pH 7)

Temp (°C) Half live
15 30 min
20 20 min
25 15 min
30 12 min
35 8 min
  • Ozone breaks down to oxygen with a half life of about 20 minutes ( at 20 degrees Celsius in water )
  • Decomposition of 2 mol (96 g) ozone to 3 mols of oxygen requires 286 kJ energy
  • pH influences acid/base equilibriums of some compounds and also the reaction speed of ozone.
  • Decay of ozone in a basic environment is much faster than in an acid environment!
  • Materials ozone resistance chart
  • Ozone decomposition kinetics

Decontamination protocol

The standard program adopted involves increasing the ozone level over a period of 15 min to 25 ppm, maintaining this level for 10 min, at which point the humidifier was activated to produce a rapid burst of water vapor. This should result in the RH increasing to > 95% within 5 min. Following this the humidifier and generator are switched off and the catalytic converter is switched on, which results in decrease in ozone to < 1 ppm within 15 min. Hudson et al.

Device requirements

  • Hermetically sealed sterilization chamber ( when door is closed )
  • Effective minimum concentration of ozone is 25 ppm ~50 mg/m3 ( 1 ppm (O3) = 2.14 mg/m3 )
  • To meet protocol parameters for every cubic meter a generator capacity of ~200 mg/h is required.
  • Air humidity shall be controlled so that RH could be increased to > 95% rapidly. ( optional )
  • There should be a timed lock, to prevent the user to open the device while operating. ( optional )
  • There should be a clear sign indicating potentially dangerous oxidant levels inside the device.
  • Ozone levels should be either measured or predicted by its decomposition model.
  • Temperature and humidity can be easily monitored. ( supporting the level prediction - optional )

System block diagram

⚠️ Note: our open design is only recommended in emergency situations, where sufficient ventilation of the room can be provided!

The system can be either completely closed or partially open, where the ionized air can travel through the chamber into a catalyst. For the sake of simplicity we chose the closed system design with the generator and the decomposer inside the sterilization chamber.

Either way the sterilizer system consists of four separate modules that can be bought or built:

  • The generator is turning atmospheric oxygen into ozone enriched air.
  • The ozone enriched air is pumped into a sealed chamber.
  • The decomposer that helps to break down ozone to oxygen.
  • The controller is regulating the sterilization process.

Open system

Closed system

A coronal discharge method based generator consists of:

  • HEPA filter on the atmospheric air intake port ( optional )
  • A fan or blower that moves the air volume along the electrodes. ( AC or low voltage DC )
  • The electrodes or "ozone plates" provide surface for the corona discharge. ( ceramic )
  • A high voltage power supply ( output ~3.5kV / 30mA ) is required to drive the plates.
  • Exhaust port to channel the O3 enriched air into an ozone resistant tube.

For ultraviolet radiation based ozone production: ( this method is less efficient )

  • Powerful light emitter with a spectral peak of 185 nm.
  • A fan or blower that moves the air.
  • The chamber has to be somewhat airtight as the ozone is harmful to humans as well.
  • Ozone, temperature and humidity sensors has to be placed inside.
  • Ideally an electric lock should block the door.

Ozone can be reduced to oxygen thermally or catalytically or by radiation.

In case of a closed system:

  • The chamber has to hold light sources at 254 nm wavelength with sufficient power to disrupt the instable bonds in ozone molecules.

In case of an open system:

  • The ozone enriched air from the chamber has to be able to enter the decomposer.
  • In case of catalytic destruction the decomposer could be a bucket of tap water.
  • If the pH can be controlled, basic environment is preferable. ( Pka = 10,3 )
  • Thermal decomposition is effective if the air flows through a tube heated to 250 C ( HL = 1.5s )

Otherwise a controller could consist of:

  • MCU chip or board ( general programmable micro controller board e.g. Nucleo or Arduino )
  • Temperature sensor to support the decomposition model and compensate for ozone readings.
  • Humidity sensor to check protocol requirements are met and compensate for ozone readings.
  • Ozone sensor as control feedback and to check protocol requirements.
  • Optional user interface ( Small OLED display and rotary encoder ) for setting chamber parameters
  • Single button to start / stop sterilization process.
  • Process indicator ( e.g. large LED )
  • Interfaces to control the ozone generator, humidifier and lock. ( e.g. solid state relays )

Build instructions

Closed system assembly

  1. Modify a refrigerator
  2. Buy an off the shelf generator module
  3. Buy an off the shelf ozone controller

How to build an Ozone Generator

How to build and Ozone Sensor

Open ozone sensor

How to build and Ozone Monitor

Open ozone monitor

How to build an Ozone Controller

Open ozone controller